{"id":23274,"date":"2026-11-20T10:06:19","date_gmt":"2026-11-20T02:06:19","guid":{"rendered":"https:\/\/www.wehopower.com\/news\/dmm-nplc-line-rejection-coverage\/"},"modified":"2026-11-20T10:06:19","modified_gmt":"2026-11-20T02:06:19","slug":"dmm-nplc-line-rejection-coverage","status":"publish","type":"news","link":"https:\/\/www.wehopower.com\/fr\/news\/dmm-nplc-line-rejection-coverage\/","title":{"rendered":"DMM NPLC Settings: Separate Line Rejection from Event Capture"},"content":{"rendered":"<div class=\"weho-blog\">\n<div class=\"key-takeaways\"><strong>Key Engineering Takeaways<\/strong><\/p>\n<ul>\n<li>NPLC sets integration duration relative to the power-line period for supported DMM measurement functions.<\/li>\n<li>A whole-cycle aperture can reject synchronous line interference while also averaging a real short-duration event.<\/li>\n<li>Integration duration and the interval between readings are different quantities; identify any unobserved interval.<\/li>\n<li>Choose separate evidence for a steady DC value and a brief voltage dip instead of treating one setting as universally best.<\/li>\n<\/ul>\n<\/div>\n<figure class=\"article-figure hero\" data-asset=\"H16-hero\"><a class=\"product-image-link\" href=\"https:\/\/www.wehopower.com\/fr\/product\/lrs-450-switching-power-supply-450w-12v-15v-24v-36v-48v-power-supply-ip20\/\" aria-label=\"LRS-450-12 details\"><img fetchpriority=\"high\" src=\"https:\/\/www.wehopower.com\/wp-content\/uploads\/2026\/11\/b3e5b862f994c866-lrs-450-12-application.avif\" alt=\"LRS-450-12 product installed in an industrial application scene with original housing and label details.\" width=\"1200\" height=\"675\" loading=\"eager\" decoding=\"async\" title=\"\"><\/a><figcaption>LRS-450-12 in an application illustration, not a customer installation photo. Scene created with AI assistance. The image does not establish a multimeter acquisition window or continuous event capture. <a href=\"https:\/\/www.wehopower.com\/wp-content\/uploads\/2026\/11\/08edf769c7ecf0a5-application-master.png\" target=\"_blank\" rel=\"noopener\">View original-resolution image<\/a><\/figcaption><\/figure>\n<nav class=\"table-of-contents\" aria-label=\"Article contents\">\n<details open>\n<summary>On this page<\/summary>\n<ol>\n<li><a href=\"#choose-the-setting-from-the-measurement-objective\">Choose the setting from the measurement objective<\/a><\/li>\n<li><a href=\"#convert-nplc-into-an-aperture-time\">Convert NPLC into an aperture time<\/a><\/li>\n<li><a href=\"#understand-what-whole-cycle-averaging-rejects\">Understand what whole-cycle averaging rejects<\/a><\/li>\n<li><a href=\"#work-a-hypothetical-short-dip-example\">Work a hypothetical short-dip example<\/a><\/li>\n<li><a href=\"#compare-settings-without-confusing-resolution-and-coverage\">Compare settings without confusing resolution and coverage<\/a><\/li>\n<li><a href=\"#verify-the-acquisition-before-accepting-the-supply-result\">Verify the acquisition before accepting the supply result<\/a><\/li>\n<li><a href=\"#send-the-instrument-mode-with-the-weho-enquiry\">Send the instrument mode with the WEHO enquiry<\/a><\/li>\n<li><a href=\"#engineering-faq\">Engineering FAQ<\/a><\/li>\n<\/ol>\n<\/details>\n<\/nav>\n<div id=\"ez-toc-container\" class=\"ez-toc-v2_0_89 counter-hierarchy ez-toc-counter ez-toc-grey ez-toc-container-direction\">\n<div class=\"ez-toc-title-container\">\n<p class=\"ez-toc-title\" style=\"cursor:inherit\">Table des mati\u00e8res<\/p>\n<span class=\"ez-toc-title-toggle\"><a href=\"#\" class=\"ez-toc-pull-right ez-toc-btn ez-toc-btn-xs ez-toc-btn-default ez-toc-toggle\" aria-label=\"Toggle Table of Content\"><span class=\"ez-toc-js-icon-con\"><span class=\"\"><span class=\"eztoc-hide\" style=\"display:none;\">Basculer<\/span><span class=\"ez-toc-icon-toggle-span\"><svg style=\"fill: #999;color:#999\" xmlns=\"http:\/\/www.w3.org\/2000\/svg\" class=\"list-377408\" width=\"20px\" height=\"20px\" viewbox=\"0 0 24 24\" fill=\"none\"><path d=\"M6 6H4v2h2V6zm14 0H8v2h12V6zM4 11h2v2H4v-2zm16 0H8v2h12v-2zM4 16h2v2H4v-2zm16 0H8v2h12v-2z\" fill=\"currentColor\"><\/path><\/svg><svg style=\"fill: #999;color:#999\" class=\"arrow-unsorted-368013\" xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"10px\" height=\"10px\" viewbox=\"0 0 24 24\" version=\"1.2\" baseprofile=\"tiny\"><path d=\"M18.2 9.3l-6.2-6.3-6.2 6.3c-.2.2-.3.4-.3.7s.1.5.3.7c.2.2.4.3.7.3h11c.3 0 .5-.1.7-.3.2-.2.3-.5.3-.7s-.1-.5-.3-.7zM5.8 14.7l6.2 6.3 6.2-6.3c.2-.2.3-.5.3-.7s-.1-.5-.3-.7c-.2-.2-.4-.3-.7-.3h-11c-.3 0-.5.1-.7.3-.2.2-.3.5-.3.7s.1.5.3.7z\"\/><\/svg><\/span><\/span><\/span><\/a><\/span><\/div>\n<nav><ul class='ez-toc-list ez-toc-list-level-1' ><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-1\" href=\"https:\/\/www.wehopower.com\/fr\/news\/dmm-nplc-line-rejection-coverage\/#Choose_the_setting_from_the_measurement_objective\" >Choose the setting from the measurement objective<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-2\" href=\"https:\/\/www.wehopower.com\/fr\/news\/dmm-nplc-line-rejection-coverage\/#Convert_NPLC_into_an_aperture_time\" >Convert NPLC into an aperture time<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-3\" href=\"https:\/\/www.wehopower.com\/fr\/news\/dmm-nplc-line-rejection-coverage\/#Understand_what_whole-cycle_averaging_rejects\" >Understand what whole-cycle averaging rejects<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-4\" href=\"https:\/\/www.wehopower.com\/fr\/news\/dmm-nplc-line-rejection-coverage\/#Work_a_hypothetical_short-dip_example\" >Work a hypothetical short-dip example<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-5\" href=\"https:\/\/www.wehopower.com\/fr\/news\/dmm-nplc-line-rejection-coverage\/#Compare_settings_without_confusing_resolution_and_coverage\" >Compare settings without confusing resolution and coverage<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-6\" href=\"https:\/\/www.wehopower.com\/fr\/news\/dmm-nplc-line-rejection-coverage\/#Verify_the_acquisition_before_accepting_the_supply_result\" >Verify the acquisition before accepting the supply result<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-7\" href=\"https:\/\/www.wehopower.com\/fr\/news\/dmm-nplc-line-rejection-coverage\/#Send_the_instrument_mode_with_the_WEHO_enquiry\" >Send the instrument mode with the WEHO enquiry<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-8\" href=\"https:\/\/www.wehopower.com\/fr\/news\/dmm-nplc-line-rejection-coverage\/#Engineering_FAQ\" >Engineering FAQ<\/a><ul class='ez-toc-list-level-3' ><li class='ez-toc-heading-level-3'><a class=\"ez-toc-link ez-toc-heading-9\" href=\"https:\/\/www.wehopower.com\/fr\/news\/dmm-nplc-line-rejection-coverage\/#Is_one_NPLC_always_a_20_ms_measurement\" >Is one NPLC always a 20 ms measurement?<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-3'><a class=\"ez-toc-link ez-toc-heading-10\" href=\"https:\/\/www.wehopower.com\/fr\/news\/dmm-nplc-line-rejection-coverage\/#Does_line-frequency_rejection_remove_a_real_short_dip\" >Does line-frequency rejection remove a real short dip?<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-3'><a class=\"ez-toc-link ez-toc-heading-11\" href=\"https:\/\/www.wehopower.com\/fr\/news\/dmm-nplc-line-rejection-coverage\/#Is_the_reading_interval_equal_to_the_aperture\" >Is the reading interval equal to the aperture?<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-3'><a class=\"ez-toc-link ez-toc-heading-12\" href=\"https:\/\/www.wehopower.com\/fr\/news\/dmm-nplc-line-rejection-coverage\/#Can_a_steady_DMM_value_prove_there_was_no_undervoltage_event\" >Can a steady DMM value prove there was no undervoltage event?<\/a><\/li><\/ul><\/li><\/ul><\/nav><\/div>\n<h2 id=\"choose-the-setting-from-the-measurement-objective\"><span class=\"ez-toc-section\" id=\"Choose_the_setting_from_the_measurement_objective\"><\/span>Choose the setting from the measurement objective<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<p>A stable DMM display can be useful when verifying a steady output value. The same display can conceal a short dip that matters to a downstream circuit. Before changing a power supply, decide whether the test needs a low-noise average or a time-resolved event record.<\/p>\n<p>Le <a href=\"https:\/\/www.wehopower.com\/fr\/product\/lrs-450-switching-power-supply-450w-12v-15v-24v-36v-48v-power-supply-ip20\/\">WEHO LRS-450-12 supply reference<\/a> identifies the product context for this external measurement plan. The examples below do not describe its measured output or a particular DMM&#39;s actual timing. They illustrate what an integrating measurement can report under declared assumptions.<\/p>\n<h2 id=\"convert-nplc-into-an-aperture-time\"><span class=\"ez-toc-section\" id=\"Convert_NPLC_into_an_aperture_time\"><\/span>Convert NPLC into an aperture time<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<p>For a supported integrating measurement, an idealized aperture duration is T = NPLC\/f<sub>line<\/sub>. At 50 Hz, one power-line cycle lasts 1\/50 = 0.020 seconds, or 20 ms. At 60 Hz, one cycle lasts approximately 16.67 ms; the same NPLC number therefore represents a different duration.<\/p>\n<p>Keysight&#39;s <a href=\"https:\/\/helpfiles.keysight.com\/Standalone_BenchVueSoftware_PW_HelpFiles\/PWDMMApp\/Content\/GUI\/Instrument%20Settings\/Measurement\/Aperture%20Time%20and%20Resolution.htm\" target=\"_blank\" rel=\"noopener\">aperture-time and resolution guidance<\/a> explains the distinction between NPLC-based integration and an aperture entered in seconds. It also discusses integer-cycle integration for normal-mode line-frequency rejection. Available settings and behavior belong to the identified DMM and function, rather than every instrument with an aperture menu.<\/p>\n<p>An aperture describes the interval contributing to one reading. Trigger handling, settling, ranging, autozero, channel changes or other operations can add time outside that interval. Determine the actual acquisition sequence instead of calculating a reading rate simply as 1\/T.<\/p>\n<figure class=\"article-figure product\" data-asset=\"H16-product\"><a class=\"product-image-link\" href=\"https:\/\/www.wehopower.com\/fr\/product\/lrs-450-switching-power-supply-450w-12v-15v-24v-36v-48v-power-supply-ip20\/\" aria-label=\"LRS-450-12 details\"><img loading=\"lazy\" src=\"https:\/\/www.wehopower.com\/wp-content\/uploads\/2026\/11\/4c603bcfab5cee3d-lrs-450-12-product-reference.avif\" alt=\"LRS-450-12 product reference preserving original housing, model marking and visible connection details for comparison.\" width=\"1000\" height=\"1000\" decoding=\"async\" title=\"\"><\/a><figcaption>WEHO LRS-450-12 product identification reference, retaining the source image&#8217;s visible identity. The photograph does not establish a multimeter acquisition window or continuous event capture. <a href=\"https:\/\/www.wehopower.com\/wp-content\/uploads\/2026\/11\/8f75f4b7d70aebcb-product-original.png\" target=\"_blank\" rel=\"noopener\">View original-resolution image<\/a><\/figcaption><\/figure>\n<h2 id=\"understand-what-whole-cycle-averaging-rejects\"><span class=\"ez-toc-section\" id=\"Understand_what_whole-cycle_averaging_rejects\"><\/span>Understand what whole-cycle averaging rejects<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<p>Consider an ideal rectangular integration window applied to a sinusoidal interference component. Its amplitude response is |sin(\u03c0fT)\/(\u03c0fT)|. At exactly 50 Hz with T = 20 ms, the numerator is sin(\u03c0), so the ideal average of that component is zero regardless of its initial phase.<\/p>\n<p>That calculation is a model of averaging, not a promised rejection specification for a meter. Actual performance depends on the instrument, its selected function, frequency handling and other errors. Line interference at a different frequency, or unrelated transient content, is not removed simply because the display says \u201c1 PLC.\u201d<\/p>\n<p>For example, at 50 Hz and T = 2 ms, corresponding to an assumed 0.1 PLC setting, the ideal amplitude ratio is sin(0.1\u03c0)\/(0.1\u03c0) \u2248 0.9836. About 98.36% of that sinusoidal amplitude remains in this model. A shorter aperture may help timing, but it does not provide the whole-cycle cancellation shown by the 20 ms example.<\/p>\n<p>Keysight&#39;s <a href=\"https:\/\/www.keysight.com\/us\/en\/use-cases\/reduce-power-line-noise-in-dc-measurements.html\" target=\"_blank\" rel=\"noopener\">DC line-noise discussion<\/a> connects integration duration, local line frequency and the reported DC value. Use it to frame the noise-rejection decision. Do not interpret noise reduction as evidence that every genuine event was observed.<\/p>\n<h2 id=\"work-a-hypothetical-short-dip-example\"><span class=\"ez-toc-section\" id=\"Work_a_hypothetical_short-dip_example\"><\/span>Work a hypothetical short-dip example<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<p>Assume an ideal integrating DMM measures a steady 12.00 V level with a 20 ms rectangular aperture. One real signal event in this hypothetical model drops the level by 0.50 V for 2 ms, entirely within the aperture. The instantaneous minimum is therefore 11.50 V.<\/p>\n<p>The aperture-average reduction is 0.50 \u00d7 (2\/20) = 0.050 V. The reported average is 12.00 \u2212 0.050 = 11.95 V, despite the modeled 11.50 V minimum. Neither number is wrong: they answer different questions about the same assumed event.<\/p>\n<p>Now assume each acquisition occupies a 25 ms cycle: 20 ms integrating followed by 5 ms of otherwise unobserved overhead. In a separate timing case, a 2 ms dip falls entirely inside that 5 ms interval. This idealized acquisition has no information about that dip and reports the unchanged level for the adjacent apertures.<\/p>\n<div class=\"table-scroll\" tabindex=\"0\" role=\"region\" aria-label=\"Scrollable engineering table\">\n<table>\n<caption>Hypothetical DC-dip cases for a 20 ms aperture and 25 ms acquisition cycle<\/caption>\n<thead>\n<tr>\n<th scope=\"col\">Case<\/th>\n<th scope=\"col\">Dip placement<\/th>\n<th scope=\"col\">Model reading<\/th>\n<th scope=\"col\">What is established<\/th>\n<\/tr>\n<\/thead>\n<tbody>\n<tr>\n<td>Steady reference<\/td>\n<td>No dip<\/td>\n<td>12.00 V<\/td>\n<td>Aperture average<\/td>\n<\/tr>\n<tr>\n<td>A: event inside aperture<\/td>\n<td>2 ms at 11.50 V<\/td>\n<td>11.95 V average<\/td>\n<td>Event contribution is averaged<\/td>\n<\/tr>\n<tr>\n<td>B: event inside overhead<\/td>\n<td>2 ms outside both apertures<\/td>\n<td>12.00 V adjacent averages<\/td>\n<td>This acquisition does not observe the event<\/td>\n<\/tr>\n<tr>\n<td>Independent minimum record<\/td>\n<td>Suitable event capture<\/td>\n<td>11.50 V in this signal model<\/td>\n<td>Needs its own real instrument evidence<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/div>\n<p>The 5 ms overhead is an explicit assumption, not a specification for all DMMs. Some acquisition modes use a different architecture and may offer contiguous records or separate event functions. Check the selected mode&#39;s actual timing, buffer and trigger behavior before describing a blind interval.<\/p>\n<figure class=\"article-figure technical\" data-asset=\"H16-technical\">\n<div class=\"diagram-scroll\" tabindex=\"0\" role=\"region\" aria-label=\"Scrollable engineering diagram\"><img loading=\"lazy\" src=\"https:\/\/www.wehopower.com\/wp-content\/uploads\/2026\/11\/5e259a55b4aff99f-dmm-aperture-event-coverage.svg\" alt=\"Hypothetical DMM timing distinguishes a short dip averaged inside an aperture from an unobserved event.\" width=\"1200\" height=\"675\" decoding=\"async\" title=\"\"><\/div><figcaption>Conceptual Engineering Diagram. Independent hypothetical timing cases use 20 ms integration and an assumed 5 ms overhead. They illustrate averaging and coverage, not actual DMM timing or a WEHO output event. <a href=\"https:\/\/www.wehopower.com\/wp-content\/uploads\/2026\/11\/5e259a55b4aff99f-dmm-aperture-event-coverage.svg\" target=\"_blank\" rel=\"noopener\">Open full-size diagram<\/a><\/figcaption><p class=\"figure-help\">Swipe horizontally on a small screen, or open the full-size diagram.<\/p>\n<\/figure>\n<p>For a project that requires evidence of no excursion below 11.70 V, the 11.95 V average cannot establish compliance. The hypothetical signal crosses that limit, but the average hides its depth. A suitable time-domain measurement and a defined duration criterion are needed to decide that requirement.<\/p>\n<p>The requirement itself must say whether a brief excursion is permitted, and for how long. A steady-state average requirement can legitimately use a different method from a minimum-voltage requirement. Keep the two results labelled rather than choosing whichever reading makes the equipment appear to pass.<\/p>\n<h2 id=\"compare-settings-without-confusing-resolution-and-coverage\"><span class=\"ez-toc-section\" id=\"Compare_settings_without_confusing_resolution_and_coverage\"><\/span>Compare settings without confusing resolution and coverage<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<p>A longer aperture can improve rejection of selected interference and improve resolution in applicable modes, but it combines signal behavior over a longer interval. A faster mode may expose more detail while increasing noise or changing resolution. Select and document the tradeoff against the actual question.<\/p>\n<p>Le <a href=\"https:\/\/www.tek.com\/en\/support\/faqs\/speed-how-can-i-refresh-auto-zero-values-once-beginning-test-sequence\" target=\"_blank\" rel=\"noopener\">Keithley autozero timing FAQ<\/a> provides a product-specific example of how zero-update choices affect a measurement sequence. Its commands and firmware conditions are not universal DMM settings. The useful lesson is to include the complete sequence when evaluating timing and drift.<\/p>\n<div class=\"table-scroll\" tabindex=\"0\" role=\"region\" aria-label=\"Scrollable engineering table\">\n<table>\n<caption>Record the setting that supports each verification task<\/caption>\n<thead>\n<tr>\n<th scope=\"col\">Task<\/th>\n<th scope=\"col\">Relevant choice<\/th>\n<th scope=\"col\">Verify<\/th>\n<th scope=\"col\">Retain separately<\/th>\n<\/tr>\n<\/thead>\n<tbody>\n<tr>\n<td>Steady DC value<\/td>\n<td>Applicable line-rejecting integration<\/td>\n<td>Function, line frequency and uncertainty<\/td>\n<td>Short-event behavior<\/td>\n<\/tr>\n<tr>\n<td>Brief minimum voltage<\/td>\n<td>Suitable time-domain\/event capture<\/td>\n<td>Bandwidth, time coverage and trigger<\/td>\n<td>Low-noise average<\/td>\n<\/tr>\n<tr>\n<td>Repeated readings<\/td>\n<td>Full acquisition sequence<\/td>\n<td>Aperture plus intervening operations<\/td>\n<td>Timestamp meaning<\/td>\n<\/tr>\n<tr>\n<td>Compare two setups<\/td>\n<td>Matched quantity and conditions<\/td>\n<td>Range, filtering and decision rule<\/td>\n<td>Different-mode diagnostic records<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/div>\n<h2 id=\"verify-the-acquisition-before-accepting-the-supply-result\"><span class=\"ez-toc-section\" id=\"Verify_the_acquisition_before_accepting_the_supply_result\"><\/span>Verify the acquisition before accepting the supply result<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<ol>\n<li><strong>Define the quantity.<\/strong> Record average voltage, minimum voltage or another named metric, its time basis and the project limit. Approve the method only when it measures that quantity rather than a convenient substitute.<\/li>\n<li><strong>Identify the mode.<\/strong> Record DMM model, function, range, NPLC or aperture, line-frequency setting, filtering and zero behavior. Confirm supported settings in the applicable instrument instructions.<\/li>\n<li><strong>Establish timing coverage.<\/strong> Record integration and trigger\/settling\/overhead behavior from applicable evidence or a qualified timing check. Do not label display refresh or interface transfer time as continuous signal observation.<\/li>\n<li><strong>Check a representative event.<\/strong> Under an approved test procedure, use an appropriate reference signal or independent acquisition to evaluate event depth, duration and timing sensitivity. Record whether the reported quantity remains sufficient for the intended decision.<\/li>\n<li><strong>Preserve both decisions.<\/strong> Retain steady-value and event records where both are required. State unresolved timing or uncertainty limitations instead of declaring an unobserved excursion absent.<\/li>\n<\/ol>\n<p>Le <a href=\"https:\/\/www.wehopower.com\/fr\/news\/power-supply-ripple-acceptance-test-matrix\/\">WEHO ripple acceptance matrix<\/a> offers a related framework for keeping source and receiver criteria separate. NPLC addresses another layer: how one reading is formed. Agreement on a limit does not remove the need to verify the acquisition behind the number.<\/p>\n<h2 id=\"send-the-instrument-mode-with-the-weho-enquiry\"><span class=\"ez-toc-section\" id=\"Send_the_instrument_mode_with_the_WEHO_enquiry\"><\/span>Send the instrument mode with the WEHO enquiry<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<p>Use the <a href=\"https:\/\/www.wehopower.com\/fr\/product\/lrs-450-switching-power-supply-450w-12v-15v-24v-36v-48v-power-supply-ip20\/\">LRS-450-12 measurement enquiry reference<\/a> with the input\/load state, test points, required quantity and complete instrument mode. Include the event duration and applicable acceptance rule. A stable display alone should not be presented as evidence that the supply or receiving equipment experienced no short dip.<\/p>\n<h2 id=\"engineering-faq\"><span class=\"ez-toc-section\" id=\"Engineering_FAQ\"><\/span>Engineering FAQ<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<h3 id=\"is-one-nplc-always-a-20-ms-measurement\"><span class=\"ez-toc-section\" id=\"Is_one_NPLC_always_a_20_ms_measurement\"><\/span>Is one NPLC always a 20 ms measurement?<span class=\"ez-toc-section-end\"><\/span><\/h3>\n<p>No, one NPLC is 20 ms at 50 Hz and about 16.67 ms at 60 Hz. Applicable modes and frequency handling depend on the instrument. Record the actual configuration.<\/p>\n<h3 id=\"does-line-frequency-rejection-remove-a-real-short-dip\"><span class=\"ez-toc-section\" id=\"Does_line-frequency_rejection_remove_a_real_short_dip\"><\/span>Does line-frequency rejection remove a real short dip?<span class=\"ez-toc-section-end\"><\/span><\/h3>\n<p>No, line rejection does not distinguish interference from every wanted event. Integration can average a short dip into a smaller reading change. Define whether the required quantity is an average or an event minimum.<\/p>\n<h3 id=\"is-the-reading-interval-equal-to-the-aperture\"><span class=\"ez-toc-section\" id=\"Is_the_reading_interval_equal_to_the_aperture\"><\/span>Is the reading interval equal to the aperture?<span class=\"ez-toc-section-end\"><\/span><\/h3>\n<p>No, intervening operations can make the interval longer. Check the actual acquisition mode and timestamps. Some modes have different coverage, so do not assume a fixed gap either.<\/p>\n<h3 id=\"can-a-steady-dmm-value-prove-there-was-no-undervoltage-event\"><span class=\"ez-toc-section\" id=\"Can_a_steady_DMM_value_prove_there_was_no_undervoltage_event\"><\/span>Can a steady DMM value prove there was no undervoltage event?<span class=\"ez-toc-section-end\"><\/span><\/h3>\n<p>No, a steady average cannot alone prove the absence of a brief event. Use a method with suitable timing coverage, bandwidth and uncertainty for the project&#39;s duration and minimum-voltage requirement.<\/p>\n<\/div>","protected":false},"excerpt":{"rendered":"<p>Choose DMM NPLC settings for DC line rejection, calculate aperture averaging, and send WEHO an LRS-450-12 test enquiry with the required event record.<\/p>","protected":false},"author":2,"featured_media":23256,"parent":0,"menu_order":38,"comment_status":"","ping_status":"","template":"","format":"standard","meta":{"_acf_changed":false,"_joinchat":[],"footnotes":""},"news_category":[284],"class_list":["post-23274","news","type-news","status-publish","format-standard","has-post-thumbnail","hentry","news_category-product-articles","entry"],"acf":[],"post_mailing_queue_ids":[],"_links":{"self":[{"href":"https:\/\/www.wehopower.com\/fr\/wp-json\/wp\/v2\/news\/23274","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.wehopower.com\/fr\/wp-json\/wp\/v2\/news"}],"about":[{"href":"https:\/\/www.wehopower.com\/fr\/wp-json\/wp\/v2\/types\/news"}],"author":[{"embeddable":true,"href":"https:\/\/www.wehopower.com\/fr\/wp-json\/wp\/v2\/users\/2"}],"replies":[{"embeddable":true,"href":"https:\/\/www.wehopower.com\/fr\/wp-json\/wp\/v2\/comments?post=23274"}],"version-history":[{"count":0,"href":"https:\/\/www.wehopower.com\/fr\/wp-json\/wp\/v2\/news\/23274\/revisions"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.wehopower.com\/fr\/wp-json\/wp\/v2\/media\/23256"}],"wp:attachment":[{"href":"https:\/\/www.wehopower.com\/fr\/wp-json\/wp\/v2\/media?parent=23274"}],"wp:term":[{"taxonomy":"news_category","embeddable":true,"href":"https:\/\/www.wehopower.com\/fr\/wp-json\/wp\/v2\/news_category?post=23274"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}