{"id":22258,"date":"2026-11-17T10:48:55","date_gmt":"2026-11-17T02:48:55","guid":{"rendered":"https:\/\/www.wehopower.com\/news\/rtd-insulation-leakage-parallel-error\/"},"modified":"2026-11-17T10:48:55","modified_gmt":"2026-11-17T02:48:55","slug":"rtd-insulation-leakage-parallel-error","status":"publish","type":"news","link":"https:\/\/www.wehopower.com\/ru\/news\/rtd-insulation-leakage-parallel-error\/","title":{"rendered":"RTD Insulation Leakage: Verify the Unwanted Parallel Path"},"content":{"rendered":"<div class=\"weho-blog\">\n<div class=\"key-takeaways\"><strong>Key Engineering Takeaways<\/strong><\/p>\n<ul>\n<li>RTD insulation leakage can divert measuring current and reduce the resistance presented to a readout.<\/li>\n<li>A parallel-resistance example explains the error mechanism without predicting a particular installed fault.<\/li>\n<li>Keep probe insulation evidence separate from lead compensation, curve selection and controller insulation ratings.<\/li>\n<li>Approve the complete sensing arrangement using applicable probe tests and a documented temperature comparison.<\/li>\n<\/ul>\n<\/div>\n<figure class=\"article-figure hero\" data-asset=\"H01-hero\"><a class=\"product-image-link\" href=\"https:\/\/www.wehopower.com\/ru\/product\/tx4-series-tx4-w-economic-dual-display-pid-temperature-controller\/\" aria-label=\"TX4-W details\"><img fetchpriority=\"high\" src=\"https:\/\/www.wehopower.com\/wp-content\/uploads\/2026\/11\/dc1746a7538df850-tx4-w-application.webp\" alt=\"RTD Insulation Leakage: Verify the Unwanted Parallel Path  title=\" width=\"1200\" height=\"675\" loading=\"eager\" decoding=\"async\" rtd insulation leakage: verify the unwanted parallel path title =\"\"><\/a><figcaption>TX4-W in an application illustration, not a customer installation photo. Scene created with AI assistance. The image does not establish probe insulation or installed leakage paths. <a href=\"https:\/\/www.wehopower.com\/wp-content\/uploads\/2026\/11\/c76124328a2bc272-application-master.png\" target=\"_blank\" rel=\"noopener\">View original-resolution image<\/a><\/figcaption><\/figure>\n<nav class=\"table-of-contents\" aria-label=\"Article contents\">\n<details open>\n<summary>On this page<\/summary>\n<ol>\n<li><a href=\"#investigate-a-low-reading-before-adding-an-offset\">Investigate a low reading before adding an offset<\/a><\/li>\n<li><a href=\"#separate-the-probe-s-insulation-from-the-controller-s-specification\">Separate the probe&#8217;s insulation from the controller&#8217;s specification<\/a><\/li>\n<li><a href=\"#calculate-a-hypothetical-parallel-path-example\">Calculate a hypothetical parallel-path example<\/a><\/li>\n<li><a href=\"#use-comparisons-that-distinguish-the-affected-boundary\">Use comparisons that distinguish the affected boundary<\/a><\/li>\n<li><a href=\"#build-an-insulation-aware-acceptance-record\">Build an insulation-aware acceptance record<\/a><\/li>\n<li><a href=\"#send-weho-the-probe-evidence-and-required-decision\">Send WEHO the probe evidence and required decision<\/a><\/li>\n<li><a href=\"#frequently-asked-questions\">Frequently asked questions<\/a><\/li>\n<\/ol>\n<\/details>\n<\/nav>\n<div id=\"ez-toc-container\" class=\"ez-toc-v2_0_89 counter-hierarchy ez-toc-counter ez-toc-grey ez-toc-container-direction\">\n<div class=\"ez-toc-title-container\">\n<p class=\"ez-toc-title\" style=\"cursor:inherit\">\u041e\u0433\u043b\u0430\u0432\u043b\u0435\u043d\u0438\u0435<\/p>\n<span class=\"ez-toc-title-toggle\"><a href=\"#\" class=\"ez-toc-pull-right ez-toc-btn ez-toc-btn-xs ez-toc-btn-default ez-toc-toggle\" aria-label=\"Toggle Table of Content\"><span class=\"ez-toc-js-icon-con\"><span class=\"\"><span class=\"eztoc-hide\" style=\"display:none;\">\u041f\u0435\u0440\u0435\u043a\u043b\u044e\u0447\u0430\u0442\u044c<\/span><span class=\"ez-toc-icon-toggle-span\"><svg style=\"fill: #999;color:#999\" xmlns=\"http:\/\/www.w3.org\/2000\/svg\" class=\"list-377408\" width=\"20px\" height=\"20px\" viewbox=\"0 0 24 24\" fill=\"none\"><path d=\"M6 6H4v2h2V6zm14 0H8v2h12V6zM4 11h2v2H4v-2zm16 0H8v2h12v-2zM4 16h2v2H4v-2zm16 0H8v2h12v-2z\" fill=\"currentColor\"><\/path><\/svg><svg style=\"fill: #999;color:#999\" class=\"arrow-unsorted-368013\" xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"10px\" height=\"10px\" viewbox=\"0 0 24 24\" version=\"1.2\" baseprofile=\"tiny\"><path d=\"M18.2 9.3l-6.2-6.3-6.2 6.3c-.2.2-.3.4-.3.7s.1.5.3.7c.2.2.4.3.7.3h11c.3 0 .5-.1.7-.3.2-.2.3-.5.3-.7s-.1-.5-.3-.7zM5.8 14.7l6.2 6.3 6.2-6.3c.2-.2.3-.5.3-.7s-.1-.5-.3-.7c-.2-.2-.4-.3-.7-.3h-11c-.3 0-.5.1-.7.3-.2.2-.3.5-.3.7s.1.5.3.7z\"\/><\/svg><\/span><\/span><\/span><\/a><\/span><\/div>\n<nav><ul class='ez-toc-list ez-toc-list-level-1' ><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-1\" href=\"https:\/\/www.wehopower.com\/ru\/news\/rtd-insulation-leakage-parallel-error\/#Investigate_a_low_reading_before_adding_an_offset\" >Investigate a low reading before adding an offset<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-2\" href=\"https:\/\/www.wehopower.com\/ru\/news\/rtd-insulation-leakage-parallel-error\/#Separate_the_probes_insulation_from_the_controllers_specification\" >Separate the probe&#39;s insulation from the controller&#39;s specification<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-3\" href=\"https:\/\/www.wehopower.com\/ru\/news\/rtd-insulation-leakage-parallel-error\/#Calculate_a_hypothetical_parallel-path_example\" >Calculate a hypothetical parallel-path example<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-4\" href=\"https:\/\/www.wehopower.com\/ru\/news\/rtd-insulation-leakage-parallel-error\/#Use_comparisons_that_distinguish_the_affected_boundary\" >Use comparisons that distinguish the affected boundary<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-5\" href=\"https:\/\/www.wehopower.com\/ru\/news\/rtd-insulation-leakage-parallel-error\/#Build_an_insulation-aware_acceptance_record\" >Build an insulation-aware acceptance record<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-6\" href=\"https:\/\/www.wehopower.com\/ru\/news\/rtd-insulation-leakage-parallel-error\/#Send_WEHO_the_probe_evidence_and_required_decision\" >Send WEHO the probe evidence and required decision<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-7\" href=\"https:\/\/www.wehopower.com\/ru\/news\/rtd-insulation-leakage-parallel-error\/#Frequently_asked_questions\" >Frequently asked questions<\/a><ul class='ez-toc-list-level-3' ><li class='ez-toc-heading-level-3'><a class=\"ez-toc-link ez-toc-heading-8\" href=\"https:\/\/www.wehopower.com\/ru\/news\/rtd-insulation-leakage-parallel-error\/#Can_RTD_insulation_leakage_produce_a_believable_low_temperature\" >Can RTD insulation leakage produce a believable low temperature?<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-3'><a class=\"ez-toc-link ez-toc-heading-9\" href=\"https:\/\/www.wehopower.com\/ru\/news\/rtd-insulation-leakage-parallel-error\/#Does_three-wire_compensation_correct_parallel_leakage\" >Does three-wire compensation correct parallel leakage?<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-3'><a class=\"ez-toc-link ez-toc-heading-10\" href=\"https:\/\/www.wehopower.com\/ru\/news\/rtd-insulation-leakage-parallel-error\/#Can_the_controllers_insulation-test_specification_be_used_on_the_connected_probe\" >Can the controller&#39;s insulation-test specification be used on the connected probe?<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-3'><a class=\"ez-toc-link ez-toc-heading-11\" href=\"https:\/\/www.wehopower.com\/ru\/news\/rtd-insulation-leakage-parallel-error\/#Should_an_offset_be_used_after_one_reference_point_agrees\" >Should an offset be used after one reference point agrees?<\/a><\/li><\/ul><\/li><\/ul><\/nav><\/div>\n<h2 id=\"investigate-a-low-reading-before-adding-an-offset\"><span class=\"ez-toc-section\" id=\"Investigate_a_low_reading_before_adding_an_offset\"><\/span>Investigate a low reading before adding an offset<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<p>An RTD can produce a stable, plausible temperature while an unwanted leakage path biases its resistance measurement. A display correction can conceal that symptom at one point without identifying the damaged insulation or unsuitable installation. Establish the electrical path and probe condition before releasing a corrected production setting.<\/p>\n<p>\u0410 <a href=\"https:\/\/www.wehopower.com\/ru\/product\/tx4-series-tx4-w-economic-dual-display-pid-temperature-controller\/\">WEHO TX4-W temperature controller<\/a> provides the receiving-instrument context for this investigation. The exact input configuration and probe still need their applicable documents. A front-panel photograph cannot establish the probe&#39;s insulation resistance, the leakage topology or the accuracy of an installed measurement.<\/p>\n<h2 id=\"separate-the-probe-s-insulation-from-the-controller-s-specification\"><span class=\"ez-toc-section\" id=\"Separate_the_probes_insulation_from_the_controllers_specification\"><\/span>Separate the probe&#39;s insulation from the controller&#39;s specification<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<p>An RTD readout is intended to determine the sensing element&#39;s resistance through its supported connection arrangement. An unwanted conductive path can allow some measuring current to bypass the intended resistance. The resulting error depends on where that path connects, not simply on whether a continuity buzzer sounds.<\/p>\n<p>The BIPM&#39;s <a href=\"https:\/\/www.bipm.org\/documents\/20126\/41773843\/BIPM_CCT_Guide_to_IPRTs.pdf\/b3bd8bba-3b0b-ab46-481f-29ceddf7bb31?download=true&amp;version=1.1\" target=\"_blank\" rel=\"noopener\">industrial platinum resistance thermometer guide, section 4.5<\/a> models insulation leakage as a resistance in parallel with the sensing resistance. That model lowers measured resistance and helps explain a low indicated temperature. It does not identify every possible conductor-to-sheath or conductor-to-earth circuit in an assembled machine.<\/p>\n<p>Probe construction and condition matter to this assessment. Fluke&#39;s <a href=\"https:\/\/www.fluke.com\/en-vn\/learn\/blog\/temperature-calibration\/why-did-my-temperature-senor-fail-calibration\" target=\"_blank\" rel=\"noopener\">temperature-sensor calibration failure guide<\/a> identifies low insulation resistance as a shunting mechanism and describes compromised seals, bent sheaths and moisture as possible contributors. Those are investigation leads rather than proof that a particular probe has failed in that way.<\/p>\n<p>Keep a controller&#39;s own insulation specification in a different record. The <a href=\"https:\/\/www.wehopower.com\/wp-content\/uploads\/2023\/11\/TX4.pdf\">WEHO TX4 manual<\/a> includes instrument insulation information, but that is not a probe-insulation test instruction. It does not authorize applying an insulation tester through the controller&#39;s sensor input or borrowing an instrument test voltage for an attached probe.<\/p>\n<figure class=\"article-figure product\" data-asset=\"H01-product\"><a class=\"product-image-link\" href=\"https:\/\/www.wehopower.com\/ru\/product\/tx4-series-tx4-w-economic-dual-display-pid-temperature-controller\/\" aria-label=\"TX4-W details\"><img loading=\"lazy\" src=\"https:\/\/www.wehopower.com\/wp-content\/uploads\/2026\/11\/69cdae9f05db9d05-tx4-w-product-reference.webp\" alt=\"RTD Insulation Leakage: Verify the Unwanted Parallel Path  title=\" width=\"1000\" height=\"1000\" decoding=\"async\" rtd insulation leakage: verify the unwanted parallel path title =\"\"><\/a><figcaption>WEHO TX4-W product identification reference, retaining the source image&#8217;s visible identity. The photograph does not establish probe insulation or installed leakage paths. <a href=\"https:\/\/www.wehopower.com\/wp-content\/uploads\/2026\/11\/33297cd7cf244510-product-original.png\" target=\"_blank\" rel=\"noopener\">View original-resolution image<\/a><\/figcaption><\/figure>\n<h2 id=\"calculate-a-hypothetical-parallel-path-example\"><span class=\"ez-toc-section\" id=\"Calculate_a_hypothetical_parallel-path_example\"><\/span>Calculate a hypothetical parallel-path example<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<p>Assume an ideal Pt100 element is exactly at 100\u00b0C, with R\u2080 = 100 \u03a9 and the positive-temperature reference coefficients A = 0.0039083\/\u00b0C and B = \u22120.0000005775\/\u00b0C\u00b2. NI&#39;s <a href=\"https:\/\/www.ni.com\/docs\/en-US\/bundle\/ni-dmm-c-api-ref\/page\/group____root__nidmm__functions__configuration__measurement__options__temperature_1gad5160554784f85f60e4e3625a9f5a238.html\" target=\"_blank\" rel=\"noopener\">RTD coefficient documentation<\/a> lists this reference characteristic. The calculation assumes zero lead error, zero self-heating and an ideal readout using that same curve.<\/p>\n<p>For temperatures at or above 0\u00b0C, use R(T) = 100 \u00d7 (1 + A \u00d7 T + B \u00d7 T\u00b2). At 100\u00b0C, the resulting resistance is 138.5055 \u03a9. These ideal conditions isolate one error mechanism; they are not TX4-W or probe performance specifications.<\/p>\n<p>Now introduce an assumed leakage resistance Rleak directly across the sensing element. The equivalent resistance is Rmeas = R \u00d7 Rleak \u00f7 (R + Rleak). This is a deliberately simplified two-node parallel model, not a claim that every sheath-leakage arrangement has exactly the same connection.<\/p>\n<p>With Rleak = 10,000 \u03a9, Rmeas = 138.5055 \u00d7 10,000 \u00f7 10,138.5055 = 136.6133 \u03a9. Inverting the same Pt100 reference curve gives approximately 95.015\u00b0C. The modeled error is therefore 95.015 \u2212 100 = \u22124.985\u00b0C, despite a perfectly stable true element temperature.<\/p>\n<figure class=\"article-figure technical\" data-asset=\"H01-technical\">\n<div class=\"diagram-scroll\" tabindex=\"0\" role=\"region\" aria-label=\"Scrollable engineering diagram\"><img loading=\"lazy\" src=\"https:\/\/www.wehopower.com\/wp-content\/uploads\/2026\/11\/31f6efc4b1ffb5b2-rtd-insulation-parallel-path.svg\" alt=\"RTD Insulation Leakage: Verify the Unwanted Parallel Path  title=\" width=\"1200\" height=\"675\" decoding=\"async\" rtd insulation leakage: verify the unwanted parallel path title =\"\"><\/div><figcaption>Conceptual Engineering Diagram. An ideal Pt100 at100\u00b0C is shunted by hypothetical leakage. The10k\u03a9 case gives136.6133\u03a9 and95.015\u00b0C; values are not TX4\/probe measurements or insulation acceptance limits. <a href=\"https:\/\/www.wehopower.com\/wp-content\/uploads\/2026\/11\/31f6efc4b1ffb5b2-rtd-insulation-parallel-path.svg\" target=\"_blank\" rel=\"noopener\">Open full-size diagram<\/a><\/figcaption><p class=\"figure-help\">Swipe horizontally on a small screen, or open the full-size diagram.<\/p>\n<\/figure>\n<div class=\"table-scroll\" tabindex=\"0\" role=\"region\" aria-label=\"Scrollable engineering table\">\n<table>\n<caption>Ideal 100\u00b0C element with an assumed parallel leakage path; calculated values only<\/caption>\n<thead>\n<tr>\n<th scope=\"col\">Assumed leakage resistance<\/th>\n<th scope=\"col\">Equivalent resistance<\/th>\n<th scope=\"col\">Reference-curve indication<\/th>\n<th scope=\"col\">Calculated error<\/th>\n<\/tr>\n<\/thead>\n<tbody>\n<tr>\n<td>No leakage path<\/td>\n<td>138.5055 \u03a9<\/td>\n<td>100.000\u00b0C<\/td>\n<td>0.000\u00b0C<\/td>\n<\/tr>\n<tr>\n<td>100 k\u03a9<\/td>\n<td>138.3139 \u03a9<\/td>\n<td>99.495\u00b0C<\/td>\n<td>\u22120.505\u00b0C<\/td>\n<\/tr>\n<tr>\n<td>10 k\u03a9<\/td>\n<td>136.6133 \u03a9<\/td>\n<td>95.015\u00b0C<\/td>\n<td>\u22124.985\u00b0C<\/td>\n<\/tr>\n<tr>\n<td>1 k\u03a9<\/td>\n<td>121.6555 \u03a9<\/td>\n<td>55.870\u00b0C<\/td>\n<td>\u221244.130\u00b0C<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/div>\n<p>The table is not an insulation acceptance standard. The assumed leakage values were selected to make the calculation visible, not to define pass or fail for a marketed probe. Actual requirements must account for construction, temperature, connection, uncertainty allocation and the probe manufacturer&#39;s applicable test method.<\/p>\n<p>The formula also shows why one fitted offset is an unreliable default remedy. The resistance error changes with element resistance and leakage resistance, and the latter may change with temperature or moisture condition. A correction obtained under one combination cannot be assumed valid across the machine&#39;s operating range.<\/p>\n<h2 id=\"use-comparisons-that-distinguish-the-affected-boundary\"><span class=\"ez-toc-section\" id=\"Use_comparisons_that_distinguish_the_affected_boundary\"><\/span>Use comparisons that distinguish the affected boundary<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<p>A useful investigation changes one boundary while preserving the rest of the comparison. A qualified resistance simulator at the controller input can investigate the receiving path, while an isolated probe assessment investigates the probe and its leads. Neither comparison alone describes the complete installed thermal measurement.<\/p>\n<div class=\"table-scroll\" tabindex=\"0\" role=\"region\" aria-label=\"Scrollable engineering table\">\n<table>\n<caption>Evidence matrix for a hypothetical low-reading investigation<\/caption>\n<thead>\n<tr>\n<th scope=\"col\">Controlled comparison<\/th>\n<th scope=\"col\">Record<\/th>\n<th scope=\"col\">What it can support<\/th>\n<th scope=\"col\">What remains open<\/th>\n<\/tr>\n<\/thead>\n<tbody>\n<tr>\n<td>Supported resistance simulator at receiver<\/td>\n<td>Input mode, value and indicated result<\/td>\n<td>Receiving-path behavior under stated conditions<\/td>\n<td>Probe insulation and process temperature<\/td>\n<\/tr>\n<tr>\n<td>Isolated probe insulation assessment<\/td>\n<td>Applicable method, temperature and result<\/td>\n<td>Probe\/lead insulation condition within test scope<\/td>\n<td>Controller conversion and installed gradients<\/td>\n<\/tr>\n<tr>\n<td>Installed reference comparison<\/td>\n<td>Probe location, stabilized reference and indication<\/td>\n<td>Agreement of that assembled arrangement<\/td>\n<td>Unrepresented temperatures or moisture states<\/td>\n<\/tr>\n<tr>\n<td>Post-change repeat<\/td>\n<td>Exact changed part and repeated conditions<\/td>\n<td>Effect of the documented correction<\/td>\n<td>Unchanged assumptions outside the test matrix<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/div>\n<p>Keep the intended electrical test separate from a thermal calibration. A satisfactory insulation result does not establish the correct Pt100 curve or immersion conditions. Conversely, agreement at one temperature does not prove that every insulation path remains suitable throughout the process cycle.<\/p>\n<p>The existing <a href=\"https:\/\/www.wehopower.com\/ru\/news\/three-wire-pt100-lead-resistance-imbalance\/\">WEHO three-wire Pt100 lead review<\/a> addresses another boundary: compensation for conductor resistance and imbalance. A three-wire connection does not remove an arbitrary parallel leakage path. Preserve both checks when both mechanisms are plausible.<\/p>\n<h2 id=\"build-an-insulation-aware-acceptance-record\"><span class=\"ez-toc-section\" id=\"Build_an_insulation-aware_acceptance_record\"><\/span>Build an insulation-aware acceptance record<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<ol>\n<li><strong>Identify the sensing assembly.<\/strong> Record the probe part number, element curve, lead arrangement, sheath, connector and exact controller configuration. Accept the definition when the electrical boundaries and applicable documents can be traced without guessing from wire colour.<\/li>\n<li><strong>Locate the suspected path.<\/strong> Document moisture exposure, transition-junction conditions and observed damage without assigning a cause prematurely. Use the probe supplier&#39;s approved investigation method; retain unresolved path details rather than substituting the ideal parallel model for evidence.<\/li>\n<li><strong>Plan the isolated tests.<\/strong> Have qualified personnel follow applicable equipment and probe instructions, including disconnection and energy control. Record the permitted test conditions and connections; never treat a controller insulation entry as permission to test through its input electronics.<\/li>\n<li><strong>Compare the relevant conditions.<\/strong> Record receiver checks, probe assessments and stabilized reference comparisons in separate rows. Judge each result against its documented requirement and uncertainty treatment, including the temperatures and exposure states the project actually needs.<\/li>\n<li><strong>Release the corrected assembly.<\/strong> Retain replacement identity, installation changes and repeated results with the configuration. Approve only the demonstrated scope; a temporary recovery after drying or handling is not automatically a permanent repair or a supplier-approved process.<\/li>\n<\/ol>\n<h2 id=\"send-weho-the-probe-evidence-and-required-decision\"><span class=\"ez-toc-section\" id=\"Send_WEHO_the_probe_evidence_and_required_decision\"><\/span>Send WEHO the probe evidence and required decision<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<p>For a <a href=\"https:\/\/www.wehopower.com\/ru\/product\/tx4-series-tx4-w-economic-dual-display-pid-temperature-controller\/\">TX4-W input review<\/a>, provide the full controller label, probe specification, operating range and connection drawing. Include the reference-comparison record and any applicable isolated probe-test report. State whether the open question concerns receiver configuration, probe condition or complete installed accuracy so the evidence request addresses the actual boundary.<\/p>\n<h2 id=\"frequently-asked-questions\"><span class=\"ez-toc-section\" id=\"Frequently_asked_questions\"><\/span>Frequently asked questions<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<h3 id=\"can-rtd-insulation-leakage-produce-a-believable-low-temperature\"><span class=\"ez-toc-section\" id=\"Can_RTD_insulation_leakage_produce_a_believable_low_temperature\"><\/span>Can RTD insulation leakage produce a believable low temperature?<span class=\"ez-toc-section-end\"><\/span><\/h3>\n<p>Yes, a suitable leakage path can lower the resistance presented to the readout. The calculated indication may remain within the configured input range. A plausible display therefore does not exclude an insulation problem.<\/p>\n<h3 id=\"does-three-wire-compensation-correct-parallel-leakage\"><span class=\"ez-toc-section\" id=\"Does_three-wire_compensation_correct_parallel_leakage\"><\/span>Does three-wire compensation correct parallel leakage?<span class=\"ez-toc-section-end\"><\/span><\/h3>\n<p>No, ordinary lead compensation does not correct an arbitrary parallel leakage path. Lead resistance and insulation leakage are different circuit effects. Confirm the actual connection and investigate each mechanism with an appropriate method.<\/p>\n<h3 id=\"can-the-controller-s-insulation-test-specification-be-used-on-the-connected-probe\"><span class=\"ez-toc-section\" id=\"Can_the_controllers_insulation-test_specification_be_used_on_the_connected_probe\"><\/span>Can the controller&#39;s insulation-test specification be used on the connected probe?<span class=\"ez-toc-section-end\"><\/span><\/h3>\n<p>No, the controller&#39;s specification is not permission to test an attached sensing assembly through its input. Use the applicable probe and instrument instructions. Record the required isolation and permitted connections before any test.<\/p>\n<h3 id=\"should-an-offset-be-used-after-one-reference-point-agrees\"><span class=\"ez-toc-section\" id=\"Should_an_offset_be_used_after_one_reference_point_agrees\"><\/span>Should an offset be used after one reference point agrees?<span class=\"ez-toc-section-end\"><\/span><\/h3>\n<p>Only use a correction when its cause and valid operating scope have been established. Leakage-related error can vary with temperature and condition. Verify the accepted arrangement across the required conditions before retaining a production correction.<\/p>\n<\/div>","protected":false},"excerpt":{"rendered":"<p>Calculate RTD leakage error and separate probe insulation from controller checks, then send WEHO your TX4-W sensing record before approving a correction.<\/p>","protected":false},"author":2,"featured_media":22218,"parent":0,"menu_order":20,"comment_status":"","ping_status":"","template":"","format":"standard","meta":{"_acf_changed":false,"_joinchat":[],"footnotes":""},"news_category":[284],"class_list":["post-22258","news","type-news","status-publish","format-standard","has-post-thumbnail","hentry","news_category-product-articles","entry"],"acf":[],"post_mailing_queue_ids":[],"_links":{"self":[{"href":"https:\/\/www.wehopower.com\/ru\/wp-json\/wp\/v2\/news\/22258","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.wehopower.com\/ru\/wp-json\/wp\/v2\/news"}],"about":[{"href":"https:\/\/www.wehopower.com\/ru\/wp-json\/wp\/v2\/types\/news"}],"author":[{"embeddable":true,"href":"https:\/\/www.wehopower.com\/ru\/wp-json\/wp\/v2\/users\/2"}],"replies":[{"embeddable":true,"href":"https:\/\/www.wehopower.com\/ru\/wp-json\/wp\/v2\/comments?post=22258"}],"version-history":[{"count":0,"href":"https:\/\/www.wehopower.com\/ru\/wp-json\/wp\/v2\/news\/22258\/revisions"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.wehopower.com\/ru\/wp-json\/wp\/v2\/media\/22218"}],"wp:attachment":[{"href":"https:\/\/www.wehopower.com\/ru\/wp-json\/wp\/v2\/media?parent=22258"}],"wp:term":[{"taxonomy":"news_category","embeddable":true,"href":"https:\/\/www.wehopower.com\/ru\/wp-json\/wp\/v2\/news_category?post=22258"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}