{"id":23830,"date":"2026-11-24T10:18:24","date_gmt":"2026-11-24T02:18:24","guid":{"rendered":"https:\/\/www.wehopower.com\/news\/probe-loading-rail-monitor-node\/"},"modified":"2026-11-24T10:18:24","modified_gmt":"2026-11-24T02:18:24","slug":"probe-loading-rail-monitor-node","status":"publish","type":"news","link":"https:\/\/www.wehopower.com\/vi\/news\/probe-loading-rail-monitor-node\/","title":{"rendered":"Oscilloscope Probe Loading: Check the Rail Monitor Test Point"},"content":{"rendered":"<div class=\"weho-l40-article\" lang=\"en-US\">\n<p>Check the node&#39;s impedance and the probe&#39;s input model before trusting a rail-monitor waveform. A probe can change the circuit it measures through resistance and capacitance. This matters especially at a high-impedance receiver monitor node; it should not be generalized to every low-impedance supply terminal.<\/p>\n<div class=\"key-takeaways\"><strong>Key Engineering Takeaways:<\/strong><\/p>\n<ul>\n<li>Identify the physical node and its source impedance.<\/li>\n<li>Use the probe&#39;s relevant input resistance and capacitance.<\/li>\n<li>Distinguish circuit loading from display or compensation errors.<\/li>\n<li>Record a controlled comparison when connecting the probe changes behavior.<\/li>\n<\/ul>\n<\/div>\n<figure class=\"article-figure hero\" data-asset=\"L10-hero\"><a class=\"product-image-link\" href=\"https:\/\/www.wehopower.com\/vi\/product\/lrs-series-400w-ac-to-dc-single-output-switching-power-supply\/\" aria-label=\"LRS-400-24 product details\"><img fetchpriority=\"high\" src=\"https:\/\/www.wehopower.com\/wp-content\/uploads\/2026\/11\/probe-loading-rail-monitor-node-hero.avif\" width=\"1200\" height=\"675\" alt=\"LRS-400-24 cabinet supply with visible mounting and routed conductors for monitor node assessment\" loading=\"eager\" decoding=\"async\" title=\"\"><\/a><figcaption>AI-assisted cabinet application illustration of LRS-400-24. The cabinet supply identifies the upstream source; the external high-impedance receiver monitor node and its probe loading are not visible. This scene is not a test record or customer installation photograph. <a href=\"https:\/\/www.wehopower.com\/wp-content\/uploads\/2026\/11\/37fc252f0794f33dc1a2c1a95114567918893f6b974b250eab576bc143146887.webp\" target=\"_blank\" rel=\"noopener\">View original image<\/a><\/figcaption><\/figure>\n<nav class=\"table-of-contents\" aria-label=\"Article contents\">\n<details open>\n<summary>On this page<\/summary>\n<ol>\n<li><a href=\"#locate-the-monitored-signal-in-the-receiver-circuit\">Locate the monitored signal in the receiver circuit<\/a><\/li>\n<li><a href=\"#recognize-that-the-probe-becomes-part-of-the-network\">Recognize that the probe becomes part of the network<\/a><\/li>\n<li><a href=\"#calculate-a-hypothetical-high-impedance-node-example\">Calculate a hypothetical high-impedance-node example<\/a><\/li>\n<li><a href=\"#compare-measurement-options-by-their-physical-effect\">Compare measurement options by their physical effect<\/a><\/li>\n<li><a href=\"#validate-the-test-point-and-the-connection\">Validate the test point and the connection<\/a><\/li>\n<li><a href=\"#provide-the-circuit-boundary-with-the-source-identity\">Provide the circuit boundary with the source identity<\/a><\/li>\n<li><a href=\"#engineering-faq\">Engineering FAQ<\/a><\/li>\n<\/ol>\n<\/details>\n<\/nav>\n<div id=\"ez-toc-container\" class=\"ez-toc-v2_0_89 counter-hierarchy ez-toc-counter ez-toc-grey ez-toc-container-direction\">\n<div class=\"ez-toc-title-container\">\n<p class=\"ez-toc-title\" style=\"cursor:inherit\">M\u1ee5c l\u1ee5c<\/p>\n<span class=\"ez-toc-title-toggle\"><a href=\"#\" class=\"ez-toc-pull-right ez-toc-btn ez-toc-btn-xs ez-toc-btn-default ez-toc-toggle\" aria-label=\"Toggle Table of Content\"><span class=\"ez-toc-js-icon-con\"><span class=\"\"><span class=\"eztoc-hide\" style=\"display:none;\">Chuy\u1ec3n \u0111\u1ed5i<\/span><span class=\"ez-toc-icon-toggle-span\"><svg style=\"fill: #999;color:#999\" xmlns=\"http:\/\/www.w3.org\/2000\/svg\" class=\"list-377408\" width=\"20px\" height=\"20px\" viewbox=\"0 0 24 24\" fill=\"none\"><path d=\"M6 6H4v2h2V6zm14 0H8v2h12V6zM4 11h2v2H4v-2zm16 0H8v2h12v-2zM4 16h2v2H4v-2zm16 0H8v2h12v-2z\" fill=\"currentColor\"><\/path><\/svg><svg style=\"fill: #999;color:#999\" class=\"arrow-unsorted-368013\" xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"10px\" height=\"10px\" viewbox=\"0 0 24 24\" version=\"1.2\" baseprofile=\"tiny\"><path d=\"M18.2 9.3l-6.2-6.3-6.2 6.3c-.2.2-.3.4-.3.7s.1.5.3.7c.2.2.4.3.7.3h11c.3 0 .5-.1.7-.3.2-.2.3-.5.3-.7s-.1-.5-.3-.7zM5.8 14.7l6.2 6.3 6.2-6.3c.2-.2.3-.5.3-.7s-.1-.5-.3-.7c-.2-.2-.4-.3-.7-.3h-11c-.3 0-.5.1-.7.3-.2.2-.3.5-.3.7s.1.5.3.7z\"\/><\/svg><\/span><\/span><\/span><\/a><\/span><\/div>\n<nav><ul class='ez-toc-list ez-toc-list-level-1' ><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-1\" href=\"https:\/\/www.wehopower.com\/vi\/news\/probe-loading-rail-monitor-node\/#Locate_the_monitored_signal_in_the_receiver_circuit\" >Locate the monitored signal in the receiver circuit<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-2\" href=\"https:\/\/www.wehopower.com\/vi\/news\/probe-loading-rail-monitor-node\/#Recognize_that_the_probe_becomes_part_of_the_network\" >Recognize that the probe becomes part of the network<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-3\" href=\"https:\/\/www.wehopower.com\/vi\/news\/probe-loading-rail-monitor-node\/#Calculate_a_hypothetical_high-impedance-node_example\" >Calculate a hypothetical high-impedance-node example<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-4\" href=\"https:\/\/www.wehopower.com\/vi\/news\/probe-loading-rail-monitor-node\/#Compare_measurement_options_by_their_physical_effect\" >Compare measurement options by their physical effect<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-5\" href=\"https:\/\/www.wehopower.com\/vi\/news\/probe-loading-rail-monitor-node\/#Validate_the_test_point_and_the_connection\" >Validate the test point and the connection<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-6\" href=\"https:\/\/www.wehopower.com\/vi\/news\/probe-loading-rail-monitor-node\/#Provide_the_circuit_boundary_with_the_source_identity\" >Provide the circuit boundary with the source identity<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-2'><a class=\"ez-toc-link ez-toc-heading-7\" href=\"https:\/\/www.wehopower.com\/vi\/news\/probe-loading-rail-monitor-node\/#Engineering_FAQ\" >Engineering FAQ<\/a><ul class='ez-toc-list-level-3' ><li class='ez-toc-heading-level-3'><a class=\"ez-toc-link ez-toc-heading-8\" href=\"https:\/\/www.wehopower.com\/vi\/news\/probe-loading-rail-monitor-node\/#Does_a_compensated_passive_probe_have_zero_loading\" >Does a compensated passive probe have zero loading?<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-3'><a class=\"ez-toc-link ez-toc-heading-9\" href=\"https:\/\/www.wehopower.com\/vi\/news\/probe-loading-rail-monitor-node\/#Can_a_probe_make_an_intermittent_symptom_disappear\" >Can a probe make an intermittent symptom disappear?<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-3'><a class=\"ez-toc-link ez-toc-heading-10\" href=\"https:\/\/www.wehopower.com\/vi\/news\/probe-loading-rail-monitor-node\/#Is_the_main_supply_output_equivalent_to_a_divider_test_point\" >Is the main supply output equivalent to a divider test point?<\/a><\/li><li class='ez-toc-page-1 ez-toc-heading-level-3'><a class=\"ez-toc-link ez-toc-heading-11\" href=\"https:\/\/www.wehopower.com\/vi\/news\/probe-loading-rail-monitor-node\/#Does_the_lowest-capacitance_probe_always_provide_the_best_answer\" >Does the lowest-capacitance probe always provide the best answer?<\/a><\/li><\/ul><\/li><\/ul><\/nav><\/div>\n<h2 id=\"locate-the-monitored-signal-in-the-receiver-circuit\"><span class=\"ez-toc-section\" id=\"Locate_the_monitored_signal_in_the_receiver_circuit\"><\/span>Locate the monitored signal in the receiver circuit<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<p>An equipment rail may feed a resistor divider, diagnostic input or monitor circuit on a receiver PCB. The accessible test point can have much higher impedance than the upstream supply terminals. Connecting an oscilloscope probe there may change the waveform or even the receiver&#39;s behavior.<\/p>\n<p>For equipment using the <a href=\"https:\/\/www.wehopower.com\/vi\/product\/lrs-series-400w-ac-to-dc-single-output-switching-power-supply\/\">WEHO LRS-400-24 source reference<\/a>, document the external receiver circuit separately from the supply. This article does not recommend probing inside the enclosed unit. Its example concerns a hypothetical monitor node with an explicitly assumed source resistance.<\/p>\n<p>Start with the circuit question. Are you observing the source voltage, the divided monitor signal or a logic response derived from that signal? A change at the monitor test point does not automatically mean that the main rail changed by the same amount. Keep each signal&#39;s transfer path visible.<\/p>\n<h2 id=\"recognize-that-the-probe-becomes-part-of-the-network\"><span class=\"ez-toc-section\" id=\"Recognize_that_the_probe_becomes_part_of_the_network\"><\/span>Recognize that the probe becomes part of the network<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<p>A probe input is not an ideal open circuit. Input resistance can alter a DC divider ratio, while input capacitance can shunt high-frequency content. The relevant impedance varies with frequency, and adapters or connection leads can introduce additional behavior beyond the probe-tip specification.<\/p>\n<p><a href=\"https:\/\/www.tek.com\/en\/documents\/application-note\/how-minimize-probe-loading-low-capacitance-probes-0\" target=\"_blank\" rel=\"noopener\">Tektronix&#39;s low-capacitance probe note<\/a> explains how capacitive loading can reduce bandwidth and increase measured rise time. Apply that mechanism using the actual node and probe data rather than transferring one example&#39;s bandwidth into another circuit.<\/p>\n<p>A simplified node model uses a Thevenin voltage source and resistance feeding the probe capacitance. It can show the scale of a loading effect, but it does not describe every receiver. Active circuitry, existing capacitors, nonlinear thresholds and interconnect can require a more complete model.<\/p>\n<p>Probe compensation addresses the probe&#39;s intended transfer relationship with the instrument. Proper compensation does not remove the input capacitance physically connected to the circuit. Similarly, a mathematically corrected display does not necessarily restore circuit behavior that changed when the probe was attached.<\/p>\n<h2 id=\"calculate-a-hypothetical-high-impedance-node-example\"><span class=\"ez-toc-section\" id=\"Calculate_a_hypothetical_high-impedance-node_example\"><\/span>Calculate a hypothetical high-impedance-node example<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<p>Assume a <strong>hypothetical<\/strong> receiver monitor node has a Thevenin resistance of 20 k\u03a9. First consider a probe presenting 12 pF at the relevant frequencies. The simple RC time constant is \u03c4 = RC = 20,000 \u00d7 12 \u00d7 10\u207b\u00b9\u00b2 = <strong>240 ns<\/strong>.<\/p>\n<p>The associated single-pole corner is fc = 1\/(2\u03c0RC), approximately <strong>663 kHz<\/strong>. The model&#39;s step 10\u201390% rise time is approximately 2.2\u03c4 = <strong>528 ns<\/strong>. These values describe the assumed loaded RC network, not the complete instrument&#39;s rise time or a supply specification.<\/p>\n<p>Now assume a suitable alternative probe presents 1.5 pF. The same calculation gives \u03c4 = <strong>30 ns<\/strong>, fc \u2248 <strong>5.31 MHz<\/strong>, and a single-pole rise time of about <strong>66 ns<\/strong>. The lower capacitance reduces the modeled loading, but its suitability still depends on voltage, noise, bandwidth, connection and other actual probe limits.<\/p>\n<figure class=\"article-figure technical\" data-asset=\"L10-technical\">\n<p class=\"scroll-help\">Swipe to view the complete diagram, or open full size.<\/p>\n<div class=\"diagram-scroll\" tabindex=\"0\" role=\"region\" aria-label=\"Scrollable engineering diagram\"><img loading=\"lazy\" src=\"https:\/\/www.wehopower.com\/wp-content\/uploads\/2026\/11\/probe-loading-rail-monitor-node-technical.svg\" width=\"1200\" height=\"675\" alt=\"Two probe capacitances create different hypothetical time constants at a high impedance voltage monitor node\" decoding=\"async\" title=\"\"><\/div><figcaption>Conceptual Engineering Diagram. Original RC loading example for a 20 k\u03a9 receiver monitor node: 12 pF gives 240 ns, while 1.5 pF gives 30 ns. The model is separate from supply-terminal impedance. <a href=\"https:\/\/www.wehopower.com\/wp-content\/uploads\/2026\/11\/977478448763a6ebcb7fda2c91b2b38e1abf2f6d45e498fc31decd522951c337.svg\" target=\"_blank\" rel=\"noopener\">Open full-size diagram<\/a><\/figcaption><\/figure>\n<p>These calculations deliberately exclude existing node capacitance and probe resistance. Include them when they are significant. An existing filter capacitor can dominate the total capacitance, while a finite probe resistance can change the DC operating value as well as the dynamic response.<\/p>\n<p>The result cannot be applied directly to a bulk DC rail without its impedance model. A low-impedance output with substantial connected capacitance may respond very differently to a few picofarads. The engineering value comes from matching the model to the particular test point, not declaring every probe connection intrusive to the same degree.<\/p>\n<h2 id=\"compare-measurement-options-by-their-physical-effect\"><span class=\"ez-toc-section\" id=\"Compare_measurement_options_by_their_physical_effect\"><\/span>Compare measurement options by their physical effect<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<p>The best comparison preserves the circuit state while changing only a justified measurement parameter. If changing probes also changes grounding, adapter length, scope bandwidth and triggering, an altered trace cannot be assigned to capacitance alone.<\/p>\n<p class=\"scroll-help\">Swipe table horizontally when needed.<\/p>\n<div class=\"table-scroll\" tabindex=\"0\" role=\"region\" aria-label=\"Scrollable engineering table\">\n<table>\n<thead>\n<tr>\n<th scope=\"col\">Ch\u1ee9ng c\u1ee9 Evidence<\/th>\n<th scope=\"col\">Question answered<\/th>\n<th scope=\"col\">Boundary Ranh gi\u1edbi<\/th>\n<\/tr>\n<\/thead>\n<tbody>\n<tr>\n<td>Node schematic and impedance<\/td>\n<td>What can the probe load?<\/td>\n<td>Must match the actual operating state<\/td>\n<\/tr>\n<tr>\n<td>Probe input model<\/td>\n<td>What resistance\/capacitance is added?<\/td>\n<td>Use relevant frequency and connection<\/td>\n<\/tr>\n<tr>\n<td>Controlled probe comparison<\/td>\n<td>Does the observation change predictably?<\/td>\n<td>Other settings remain documented<\/td>\n<\/tr>\n<tr>\n<td>Receiver functional record<\/td>\n<td>Does probing alter operation?<\/td>\n<td>Trace alone may omit system behavior<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/div>\n<p><a href=\"https:\/\/www.tek.com\/en\/documents\/application-note\/how-oscilloscope-probes-affect-your-measurement\" target=\"_blank\" rel=\"noopener\">Tektronix&#39;s probe-measurement explanation<\/a> notes that loading can affect circuit operation as well as measurement. Keep that distinction in the report. A probe that suppresses the symptom may be changing the system rather than proving the original node was healthy.<\/p>\n<p>Input resistance needs an equally explicit check for a divided rail. Calculate the loaded ratio using the actual probe resistance and network. A probe selected for low capacitance can still be inappropriate if another input characteristic or rating conflicts with the circuit.<\/p>\n<h2 id=\"validate-the-test-point-and-the-connection\"><span class=\"ez-toc-section\" id=\"Validate_the_test_point_and_the_connection\"><\/span>Validate the test point and the connection<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<ol>\n<li>Identify the exact node, reference and circuit state. Obtain the receiver schematic or an applicable impedance model before assigning a loading explanation.<\/li>\n<li>Check probe and channel ratings, input characteristics and required connections. Establish that the reference connection is suitable for the circuit.<\/li>\n<li>Estimate the expected DC and dynamic effects using a model appropriate to the node. Label assumptions that lack direct component evidence.<\/li>\n<li>Repeat with an independently suitable measurement configuration while preserving other conditions. Save the full setup and any change in receiver behavior.<\/li>\n<li>Compare the result with the model and acceptance requirement. If loading remains significant, revise the method or test-point design before relying on the waveform.<\/li>\n<\/ol>\n<figure class=\"article-figure product\" data-asset=\"L10-product\"><a class=\"product-image-link\" href=\"https:\/\/www.wehopower.com\/vi\/product\/lrs-series-400w-ac-to-dc-single-output-switching-power-supply\/\" aria-label=\"LRS-400-24 product details\"><img loading=\"lazy\" src=\"https:\/\/www.wehopower.com\/wp-content\/uploads\/2026\/11\/probe-loading-rail-monitor-node-product.avif\" width=\"1000\" height=\"1000\" alt=\"WEHO LRS-400-24 product reference showing enclosure label and terminal arrangement for monitor node assessment\" decoding=\"async\" title=\"\"><\/a><figcaption>Original WEHO LRS-400-24 product photograph for identifying the supply in the review. The circuit example concerns an external test point, not an internal node of the supply. <a href=\"https:\/\/www.wehopower.com\/wp-content\/uploads\/2026\/11\/44622179442b7304287b0a0f38978f8f6d6a48cbe10ef1bb5b32fb746d4754df.jpg\" target=\"_blank\" rel=\"noopener\">View original image<\/a><\/figcaption><\/figure>\n<p>Do not improvise grounding to improve a trace. Earth-referenced scope connections can create unintended circuit paths. Use the instrument&#39;s rated method and qualified personnel where electrical conditions require them; removing protective earth is not a measurement technique.<\/p>\n<p>A buffered test point may reduce one loading problem, but the buffer becomes another circuit with its own bandwidth, noise and input\/output limits. Validate the complete path. An added component should not be described as transparent merely because its nominal input resistance is high.<\/p>\n<p>If the node is part of an acceptance fixture, preserve its probing method in the work instruction. A substituted probe or adapter can change the effective load without changing the displayed volts-per-division. Include probe identity and connection geometry in the controlled setup rather than leaving them to operator preference.<\/p>\n<p>For intermittent issues, record whether the symptom occurred before attaching the probe, after attaching it and after removing it under an approved procedure. That sequence is diagnostic context, not proof of cause by itself. Other environmental or state changes may occur at the same time.<\/p>\n<p>An independent reference instrument can add its own capacitance or resistance to the same node. Connecting two instruments simultaneously may therefore create a third configuration rather than a transparent comparison. Include every attached input in the circuit model and explain whether records were acquired together or sequentially.<\/p>\n<p>If a test-point redesign is needed, preserve the receiver&#39;s intended filtering and diagnostic behavior. Bypassing an existing filter can make a trace faster while changing the functional signal being assessed. The measurement arrangement should answer the original circuit question rather than create a more convenient but different node.<\/p>\n<h2 id=\"provide-the-circuit-boundary-with-the-source-identity\"><span class=\"ez-toc-section\" id=\"Provide_the_circuit_boundary_with_the_source_identity\"><\/span>Provide the circuit boundary with the source identity<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<p>For an assembly containing the <a href=\"https:\/\/www.wehopower.com\/vi\/product\/lrs-series-400w-ac-to-dc-single-output-switching-power-supply\/\">WEHO LRS-400-24 cabinet source<\/a>, send the source identity, external receiver schematic, test-point location, probe model and connection record. Include the load state and the specific waveform or behavior that changes when probing.<\/p>\n<p>A useful handoff distinguishes source-terminal voltage, receiver monitor response and measurement-induced changes. That separation lets the teams assess the relevant component and interface without turning an external monitor-node observation into an unsupported claim about the supply&#39;s internal regulation.<\/p>\n<h2 id=\"engineering-faq\"><span class=\"ez-toc-section\" id=\"Engineering_FAQ\"><\/span>Engineering FAQ<span class=\"ez-toc-section-end\"><\/span><\/h2>\n<h3 id=\"does-a-compensated-passive-probe-have-zero-loading\"><span class=\"ez-toc-section\" id=\"Does_a_compensated_passive_probe_have_zero_loading\"><\/span>Does a compensated passive probe have zero loading?<span class=\"ez-toc-section-end\"><\/span><\/h3>\n<p>Compensation does not eliminate the probe&#39;s physical input resistance and capacitance. It adjusts the intended measurement transfer relationship. Evaluate circuit loading separately using the actual node and probe characteristics.<\/p>\n<h3 id=\"can-a-probe-make-an-intermittent-symptom-disappear\"><span class=\"ez-toc-section\" id=\"Can_a_probe_make_an_intermittent_symptom_disappear\"><\/span>Can a probe make an intermittent symptom disappear?<span class=\"ez-toc-section-end\"><\/span><\/h3>\n<p>A probe can change a susceptible node enough to alter behavior. Record that observation and compare it with an applicable loading model. The change supports investigation but does not alone establish the original failure mechanism.<\/p>\n<h3 id=\"is-the-main-supply-output-equivalent-to-a-divider-test-point\"><span class=\"ez-toc-section\" id=\"Is_the_main_supply_output_equivalent_to_a_divider_test_point\"><\/span>Is the main supply output equivalent to a divider test point?<span class=\"ez-toc-section-end\"><\/span><\/h3>\n<p>The two nodes can have very different source impedance and connected capacitance. Use the actual circuit boundary. A high-impedance monitor-node example should not be assigned to low-impedance output terminals without evidence.<\/p>\n<h3 id=\"does-the-lowest-capacitance-probe-always-provide-the-best-answer\"><span class=\"ez-toc-section\" id=\"Does_the_lowest-capacitance_probe_always_provide_the_best_answer\"><\/span>Does the lowest-capacitance probe always provide the best answer?<span class=\"ez-toc-section-end\"><\/span><\/h3>\n<p>Capacitance is one selection factor. Voltage ratings, noise, bandwidth, resistance, connection method and calibration also matter. Choose a probe whose complete characteristics support the measurement requirement.<\/p>\n<\/div>","protected":false},"excerpt":{"rendered":"<p>Check probe capacitance and node impedance before trusting a rail-monitor waveform. Send WEHO the source model, receiver circuit and measurement settings.<\/p>","protected":false},"author":2,"featured_media":23812,"parent":0,"menu_order":0,"comment_status":"","ping_status":"","template":"","format":"standard","meta":{"_acf_changed":false,"_joinchat":[],"footnotes":""},"news_category":[284],"class_list":["post-23830","news","type-news","status-publish","format-standard","has-post-thumbnail","hentry","news_category-product-articles","entry"],"acf":[],"post_mailing_queue_ids":[],"_links":{"self":[{"href":"https:\/\/www.wehopower.com\/vi\/wp-json\/wp\/v2\/news\/23830","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.wehopower.com\/vi\/wp-json\/wp\/v2\/news"}],"about":[{"href":"https:\/\/www.wehopower.com\/vi\/wp-json\/wp\/v2\/types\/news"}],"author":[{"embeddable":true,"href":"https:\/\/www.wehopower.com\/vi\/wp-json\/wp\/v2\/users\/2"}],"replies":[{"embeddable":true,"href":"https:\/\/www.wehopower.com\/vi\/wp-json\/wp\/v2\/comments?post=23830"}],"version-history":[{"count":0,"href":"https:\/\/www.wehopower.com\/vi\/wp-json\/wp\/v2\/news\/23830\/revisions"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.wehopower.com\/vi\/wp-json\/wp\/v2\/media\/23812"}],"wp:attachment":[{"href":"https:\/\/www.wehopower.com\/vi\/wp-json\/wp\/v2\/media?parent=23830"}],"wp:term":[{"taxonomy":"news_category","embeddable":true,"href":"https:\/\/www.wehopower.com\/vi\/wp-json\/wp\/v2\/news_category?post=23830"}],"curies":[{"name":"wp c\u1ee7a t\u00f4i","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}